Temperature & Pressure Instrument

CALIBRATION INSTRUMENT

SPECIFICATION

PRESSURE CALIBRATOR

FLUKE 719-30G

fluke_719~1

  • Pressure Measurement
  • Range : -12 to 30 PSI / -0.8 to 2 bar
  • Resolution : 0.001 PSI, 0.0001 bar
  • Akurasi : ± 0.025 %

GAUGE PRESSURE MODULE

FLUKE 750P27

f2

  • Pressure Measurement
  • Range : 0 to 300 PSI / 0 to 20 bar
  • Akurasi : ± 0.0175 %

MULTI FUNCTION CALIBRATOR

FLUKE 724

f3

Measurement Accuracy

Voltage DC 30 V 0.02% + 2 counts (upper display)
20 V 0.02% + 2 counts (lower display)
100 mV 0.02% + 2 counts
-10 mV to 75 mV 0.025% + 1 count (via TC connector)
Current DC 24.000 mA 0.02% + 2 counts
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)

Source Accuracy

Voltage DC 100.00 mV 0.02% + 2 counts
10.000 V 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)
Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)

RTDs and Thermocouples

Measure Accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.8°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U:  0.1°C, 0.1°F
B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source Accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.4°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C
B, R, S: 1°C
XK 0.6°C
BP 1.2°C